After the heat-up phase has been completed, the control loop actual values are indicated.
Behavior of keys during adjustments
Pressing a key can increase or reduce a value gradually:
• Keeping the “–“ or “+” key depressed switches to rapid scan mode,
• after the key has been kept depressed for more than 3 seconds, the quick run speed increases.
Setting the temperature set value
1. Press the
Temperature
key.
• The Temperature menu (
) is displayed.
Figure 6-7.
Temperature display field and temperature selection menu
2. To exit the Temperature menu:
• Press the
End
key [3].
3. To set the temperature set value:
• Press the
Set value
key.
Note
CO
2
gas supply:
During the 5-minute heat-up phase of the O
2
control circuit, the CO
2
supply to the workspace and
monitoring of the CO
2
control circuit are suspended.
Note
Saving the settings:
To save changed values, press the
Enter
key.
Resetting the settings:
Unless a user action (contact with the pressure-sensitive areas and keys) occurs within 30 seconds,
the system automatically exits the menu and the most recently confirmed setting defaults.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...