4-1
4
Start-up
Adapting the unit to the ambient atmosphere
Preparing the workspace
Upon delivery, the CO
2
incubator is not in a sterile state. Before the initial start-up, the device must be
decontaminated.
Before the decontamination is performed, the following workspace components must be cleaned:
• Support rail
• Shelf support
Table of Contents
•
“Adapting the unit to the ambient atmosphere“
•
•
“Installing the “MAX” fill level indicator and the pre-filter“
•
•
•
“Mount the split inserts (optional) into the STERI-CYCLE i250“
•
•
•
“Connecting the USB interface:“
•
“Connecting the alarm contact:“
CAUTION
Adapting the unit to the ambient atmosphere!
Prior to commissioning allow sufficient time the unit to adapt to the ambient atmosphere.
• Prior to powering up, place the unit for approx. 2 hours in the operating room at the expected
prevailing operating room temperature.
• Open device doors.