6
Handling and control
User configuration
6-34
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Figure 6-38.
Low humidity setting
1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
The system returns to the
Options
menu.
Upon return to the main menu the
Low humidity
icon appears.
Gas tight screen setting
As the aperture cross-sections are smaller when accessing the samples, devices equipped with the
optional gas-tight screen achieve shorter recovery times for the incubation parameters:
• Workspace temperature,
• CO
2
concentration,
• O
2
concentration,
• relative humidity.
After conversion of the device, the device control system must be adjusted to the gas-tight screen
option.
Gas tight screen setting
1. Press the
Menu
key.
2. Select the menu command
Options
.
Note
Low humidity:
The enabling/disabling of the Low humidity function is entered into the event list.
Note
Malfunction:
The conversion to gas-tight screen causes a change of the control parameters.
Failure to set the gas-tight screen function in accordance with the actually installed door version
may impair the incubation performance.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...