6
Handling and control
Measures after power outage
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
6-51
*1)
This error appears on the display only; it is not entered in the error log.
*2)
When this error occurs, a special control mode is activated to protect the samples. An icon appears
to indicate this mode.
*3)
waiting time before error message:
- 45 min. after a door was opened,
- 159 min. after a setpoint change.
*4)
This setting can be changed by field service.
0...100%
O
2
Sensor breakage
Measured value exceeds
accepted limit
Request service
X
X
X
Actual value above
Actual value > Set value
+ 1%
*4)
Check the gas supply
source. Reduce
primary pressure to
1 bar (14.5 psi) max.
X
X
X
Actual value below
Actual value < Set value
- 1%
*4)
Check the gas supply
source. Replace gas
cylinder. Increase
primary pressure to 1
bar (14.5 psi) max.
Check feeding line.
X
X
X
Error communication
Sensor not
communicating with
the main board
Request service
X
X
X
Error: Gas cylinder
changeover switch
Gas cylinder
changeover switch not
communicating with
the main board
automatic
X
X
X
No gas
Both O
2
gas cylinders
are empty
Swap at least one O
2
cylinder.
X
X
X
Cylinder A empty
Gas cylinder A is empty Replace gas cylinder A
Cylinder B empty
Gas cylinder B is empty Replace gas cylinder B
rH
No water
Not enough water in
water reservoir.
Refill water or, if dry
operation is desired,
de-activate the water
level sensor. In case
the failure message
occurs again after
filling, troubleshoot
the device according
to Section
. If the
error persists, request
service.
X
X
X
Control
loop
Failure message
Cause
Remedy
Alarm
relay
Audible
alarm
Log
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...