6
Handling and control
Measures after power outage
6-48
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Troubleshooting
The error tables indicate source of error, cause of error, and possible corrections.
For any communication with the Technical Service of Thermo Fisher Scientific, please have the device
data ready.
Control
loop
Failure message
Cause
Remedy
Alarm
relay
Audible
alarm
Log
System
Device door open too
long
Device door has been
open for longer than 10
minutes
Close device door
X
X
X
Error: Display
Display not
communicating with
the main board
*1)
Reset unit. If the
error occurs
repeatedly, call
service.
X
X
X
Error: EEPROM main
board
EEPROM on main
board is defective
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
Error: data logger
Error while writing to
the memory of the data
logger. Incubator is still
operative.
Reset unit. If the error
occurs repeatedly, call
service.
Error: steri-run
Error in Steri-Run
routine
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
Power down during
steri-run
Power failure during
Steri-Run routine
Restart unit and
re-launch Steri-Run.
X
X
X
Error: auto-start
Error in auto-start
routine
Re-launch auto-start.
If the error occurs
repeatedly, call ser-
vice.
X
X
X
Error: ADC
Measurement of
reference resistance out
of tolerance
Reset unit. If the error
occurs repeatedly, call
service.
x
x
x
Error: Fan
Actual value of the fan
is out of tolerance.
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
IR sensor changed
New serial number
detected
Acknowledge the
alarm
X
X
X
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...