5
Installation
Page 56
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
A common ground for all system modules will avoid ground loops, which can
create erratic results (for example, high baseline noise).
CAUTION—Electric Shock or Damage to the Device
Connecting the device to a line voltage higher or lower than
specified could result in personal injury or damage to the device.
Therefore, connect the device to the specified line voltage only.
Never connect the device to a power socket that is shared with other
equipment (for example, multiple sockets).
Do not use defective multiple socket or extensions cords, as they
could result in personal injury or damage to the device.
After the power to the device is turned off, the device is still
energized as long as the power cord is connected. Repair work on
the device while the device is connected to power could lead to
personal injury. Therefore, always unplug the power cord before
starting repair work inside the device. If you were instructed to
remove any covers or panels, do not connect the power cord to the
device while the cover or panels are removed.
Summary of Contents for ISQ EC
Page 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Page 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Page 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Page 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...