8
Troubleshooting
Page 210
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
8.3
Troubleshooting Leaky Source Gases
If the ISQ EC or ISQ EM is having problems with leaky source gases, try
tightening the spray chamber pins.
1)
Shut down and vent the instrument as described in Section
2)
Remove the source housing as described in Section
3)
Check if the spray chamber pins are tight. If they are loose, use a T10
screwdriver to tighten the spray chamber guide pins as shown in Figure
122.
Figure 122: Tightening spray chamber pins