5
Installation
Page 94
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
3)
Add pump oil into the reservoir. Fill the pump a little over the halfway mark on the sight
glass. See Figure 39.
Figure 39: Adding oil to the foreline pump
4)
Close and tighten the reservoir cap. See Figure 40.
NOTICE—After tuning on the pump, the oil level might drop. If the oil
level goes below halfway on the sight glass, vent the mass spectrometer
and add oil.