7
Maintenance and Service
Page 196
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Figure 115: Removing Q00 from cage assembly
13)
Q00 and L0 should now be separate from the cage assembly. See
Figure 116.
Figure 116: Q00 and L0
14)
Using a small Allen key, unscrew the PEEK piece between the jumper
wires. Remove the piece. See Figure 117