5.4
Unpacking the Instrument
It is crucial to remove the black foam insert that is wedged between the
power supply module and the manifold. The foam insert keeps the glass top
of the manifold in-place during shipping. In order to remove the shipping
foam, please follow the instructions below:
1)
Be sure the instrument is unplugged.
2)
Using the provided tool kit, remove the top panel and right-hand
side panel of the instrument.
3)
Unscrew the power supply.
4)
Lift up the power supply module as shown in Figure 3.
5)
Remove the black foam insert.
6)
Swing the power supply module back in place and secure it with the
screw.
7)
Replace the side and top panels.
Figure 3: Power supply module and location of manifold foam insert