Operation
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 149
Figure 82: Method Editor—Basic Mode
Parameters and descriptions of the method editor basic mode settings are
in .
Table 12: Scan parameters: Basic Mode
Table 13: Fraction collection channels parameters: Basic Mode
Parameter
Description
Scan name
The name of the compound being scanned
Mass list or range (amu)
The masses the system will scan. Full-scan mode: Start and end mass
are separated by a dash. SIM mode: Individual masses or multiple
values separated by a comma.
Ion Polarity
Indicates whether positive or negative ions are generated
Source CID voltage
Accelerates ions into the background gas to aid in desolvation. Can
increase sensitivity or reduce noise for some analytes. Range: 0-100
V, default: 0 V (no source CID)
Transfer scans
Select button to transfer scan information typed in Basic Mode to
Scan Mode
Parameter
Description
Name
The name of the channel for fraction collection
Scan Filter
Based on the scan information entered into the Scans pane
Target Mass (min)
The minimum mass scanned within the scans range
Target Mass (max)
The maximum mass scanned within the scans range
Summary of Contents for ISQ EC
Page 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Page 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Page 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Page 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Page 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...