The
Hardware Test
screen is displayed when the self-verification test is complete. The
Hardware
Test
screen displays the functions that pass and fail.
4.
(Optional)
Tap
Export
to export the results.
5.
Tap
Close
.
Use Smart Help if any hardware functions fail (see “Request technical support with Smart Help”
).
View the last hardware test result
The self verification test checks the instrument hardware functions. You can view the results of the last
self-verification test that was run.
1.
In the home screen, tap
(Settings)
4
Maintenance and service
4
Self verification test
.
The
Self-Verification Test
screen is displayed.
2.
Tap
Hardware test
.
3.
Tap
Last test
.
The
Self-Verification Test
screen displays the functions that pass and fail.
Use Smart Help if any hardware functions fail (see “Request technical support with Smart Help”
).
Perform a peripheral test
The peripheral test takes approximately 5 minutes. Tap
(Close)
on any screen to exit the peripheral
test.
The peripheral test checks the following functions:
• Touchscreen accuracy
• Speaker
• Microphone
• Camera
• Proximity sensor
1.
In the home screen, tap
(Settings)
4
Maintenance and service
4
Self verification test
.
The
Self Verification Test
screen is displayed.
2.
Tap
Peripheral test
.
3.
Tap
Start
.
The peripheral test begins with the touch accuracy test.
Chapter 11
Configure the instrument settings
Configure maintenance and service settings
11
174
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide