Set up or remove the Auto Delta feature
Note:
The Auto Delta feature is only available in the PCR stage.
1.
In the
Method
tab, tap
Edit
4
Advanced
4
VeriFlex
™
Zones / Auto Delta
.
Note:
Use or to navigate through the method if it is not all displayed.
2.
Tap
(Edit)
in one of the PCR stages.
3.
Tap
Auto Delta settings
.
4.
Enter the starting cycle to apply Auto Delta settings.
5.
Set the
Delta temperature
slider to the position or the position, then enter the numerical
value for the difference in the temperature.
6.
Set the
Delta time
slider to the position or the position, then enter the numerical value for
the difference in the time.
7.
Tap
Done
.
A PCR step with an Auto Delta setting applied to it is denoted with an .
8.
In the
Method
tab, tap
4
Remove
4
Done
to remove the Auto Delta setting.
Note:
The Auto Delta setting is removed from the
Method
tab.
1
2
1
Indicates that this is the
Method
tab where Auto Delta settings can be removed.
2
Indicates that Auto Delta is assigned. Tap to remove the Auto Delta settings.
Chapter 4
Edit a plate before starting a run
Options to edit the method of a plate file
4
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide