Environmental requirements
Condition
Acceptable range
Installation site
Indoor use only
Electromagnetic
interference
Do not use this device in close proximity to sources of strong electromagnetic radiation (for
example, unshielded intentional RF sources). Strong electromagnetic radiation may interfere
with the proper operation of the device.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference. You may need to take measures to mitigate
the interference.
Altitude
Located between sea level and 2000 m (6500 ft.) above sea level
Humidity (instrument
and computer)
Operation: 15%–80% (noncondensing)
Temperature
(instrument and
computer)
15°C to 30°C (60°F to 85°F)
Note:
The room temperature must not fluctuate more than 2°C over a 2-hour period.
Transient category
Installation categories II
Overvoltage category Installation categories II
Vibration
The instrument is not adjacent to strong vibration sources, such as a centrifuge, pump, or
compressor. Excessive vibration will affect instrument performance.
Pollution degree
II
Install the instrument in an environment that has nonconductive pollutants such as
dust particles or wood chips. Typical environments with a Pollution Degree II rating are
laboratories and sales and commercial areas.
Liquid waste
collection
Dispose of the polymer, buffer, reagents and any liquid waste as hazardous waste in
compliance with local and national regulations.
Other conditions
Ensure the room is away from any vents that could expel particulate material on the
components.
Avoid placing the instrument and computer adjacent to heaters, cooling ducts, or in direct
sunlight.
Appendix A
Instrument specifications and layout
Environmental requirements
A
192
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide