Architecture
The following component diagram describes the communication within subcomponents of
DynaCellect
™
instrument and to the external OPC-UA Client (DeltaV
™
system).
The Instrument Server will be the main interface performing back end logic to support both eGUI and
OPC-UA Server interactions. The DeltaV
™
system will establish login authentication and communication
exchange with the OPC server that is a part of DynaCellect
™
system.
Chapter 4
OPC-UA interface
MES Interface
4
CTS
™
DynaCellect
™
Magnetic Separation System User Guide