Product specifications
Category
Property
Value
Electrical
Supply voltage
100–240 VAC ±10%, 50/60 Hz,
600 W
Phases
Single
Maximum rated input current
6 A
Fuses
2 × 10 A
Instrument
Regulatory requirements
HEPA filter meets ISO 14644-14
Installation category
II
Instrument type
Benchtop/cart unit
Instrument dimensions
495 mm (19") with pole base/
434 mm (17") without pole
base (H) × 749.5 mm (29.5")
(W) × 562 mm (22.12") (D)
Instrument weight
49.75 kg (109.7 lbs)
(50.6 kg/111.6 lbs with poles and
bag hangers)
Built-in features
8” touch screen display
(800 × 600 px)
Sound level
Operating sound level
73 dBA (measured 1 m from
instrument)
Non-operating sound level
63 dBA (measured 1 m from
instrument)
Environmental ranges
Ambient temp
15°C to 30°C
Transport temperature
-30°C to 60°C (extreme end
temperatures at maxium 72 hours)
Storage temperature
15°C to 30°C
Relative humidity
15%–80% (non-condensing)
Operating limits
Magnet
Engaged/Disengaged
Chapter 1
Product information
System overview
1
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
15
Summary of Contents for A55867