Rocker
Pump active
When the rocker has been activated (
On
), it can be programmed as shown in the following image.
Note:
When the pump is active the pump duration dictates the duration of the step and the
magnet/rocker activity will continue for the entire step duration.
The duration windows are grey, reflecting that the duration cannot currently be altered. This is because
the duration of the pump determines the step duration (42 seconds, as shown in the following image),
which means the rocker cannot have a different duration within the same step.
Chapter 3
Embedded graphical user interface (eGUI) user instructions
Manage protocols
3
62
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
Summary of Contents for A55867