MCB error list
Code
Message
Context
Description
Recovery Action
0x8106
The motor driver
encountered a fault
condition
Recipe
Manual
Motor driver fault
error
• Power cycle
instrument
• If error persists,
export service
logs and send
to tech support
for further
assistance
0x8124
Home switch not
found
Startup
Motor failed to
detect home switch
Export service logs
and send to tech
support for further
assistance
0x8125
Could not move out
of home switch
Startup
Motor failed to
move out of home
switch
Export service logs
and send to tech
support for further
assistance
0x8801
Magnet jack failed
to move up
Recipe
Manual
Magnet jack motor
up position sensor
failed to activate
after move up
command
Export service logs
and send to tech
support for further
assistance
0x8802
Magnet jack failed
to move down
Recipe
Manual
Magnet jack motor
down position
sensor failed to
activate after move
down command
Export service logs
and send to tech
support for further
assistance
0x8803
Magnet jack
position sensors are
both active
Startup
Recipe
Manual
Both magnet jack
up and down
position sensors are
activated
Export service logs
and send to tech
support for further
assistance
0x8804
Magnet jack motor
failed to home
Startup
Manual
Magnet jack motor
failed to find its
home position
Export service logs
and send to tech
support for further
assistance
0x8805
Bag lid height
motors failed to
extend
Recipe
Manual
Bag lid height motor
extend position
sensor failed to
activate after extend
command
Export service logs
and send to tech
support for further
assistance
Appendix B