2-2
Type II Water System
Thermo Scientific
Section 2
Product Information
Table 2-1.
Feed Water Requirements
1
1500 ppm NaCl feed water solution @ 25°C (77°F), @ 4.5 bar (65 psig), 15%
recovery. Flow rate and recovery will decrease with lower water temperature and
pressure. We recommend the use of a hot water mixing valve before the RO system for
water cooler than 15°C (59°F). Feed water suitability must be determined by a water
analysis.
Table 2-2.
RO System
Table 2-3.
Membrane Performance
2
2
500 ppm NaCl feed water solution @ 25°C (77°F), @ 4.5 bar (65 psig), 15%
recovery. Flow rate and recovery will decrease with lower water temperature and
pressure. The use of a hot water mixing valve before the RO system is recommended for
water cooler than 15°C (59°F).
Types
Tap
1
(Potable)
Turbidity
1.0 N.T.U. maximum
Pressure Range
2-6.9 bar (30-100psig)
Temperature Range
4-40°C (39-104°F)
Minimum Inlet Flow Requirements
100 lph
pH
3-10
TDS
600 (max. ppm as CaCO3)
Silt Density Index
< 5%
Free Chlorine
< 3 ppm
Langlier Saturation Index
< 1
Iron (total as Fe)
< 0.5 ppm
Silica
< 30 ppm
TII Water
Specifications
Model
RO Product Flow Rate
Recovery
D14031
12 lph (3.2gph), ±15%
10-20%
D14041
24 lph (6.3gph), ±15%
20-40%
Rejection
Inorganic (minimum)
>90%
Inorganic (typical)
>95%
Particles
>99%
Bacteria
>99%
Organics (>300 MW)
>99%