UP arrow: . . . . . . . . . . . . .Increments by 5 minutes
DOWN arrow: . . . . . . . . . .Decrement by 1 minute
Maximum: . . . . . . . 30 or 60 minutes depending on reservoir volume
Minimum: . . . . . . . . . . . . . . . . . . . . . . . . .1 minute
UP arrow . . . . . . . . . . . . . . . . .250ml up to 1 liter, then increments of 5L
DOWN arrow . . . Decrement by 250 ml up to 1L, then decrement by 1L
Maximum . .. . . . . . . . . . . . . .30L or 60L depending on reservoir volume
Minimum: . . . . . . . . . . . . . . . . . . . . . . . . . . .250 ml
If the water level is below 75%, the external reservoir will begin
automatically refilling in either RUN mode or STANDBY mode by the
self-contained RO system. The water level in the reservoir is determined by
internal sensors in the reservoir. Once the water level fills up to the full
level, the RO system will enter STANDBY mode.
Note
When the RO operation is automatically turned on, RO water will
be put into the reservoir. Anytime the RO purity falls below the user-
selected % rejection setting, the System Alarm LED will illuminate.
s
1. Upon power up, press any of the four buttons below the display when
the display shows LANGUAGE.”
2. Press the UP or DOWN arrow to select the language desired and press
OK.
1. Press MENU and scroll to the SET TIME screen and press OK. Using
the arrow keys, select time and press OK.
2. Scroll to the SET DATE screen and press OK. Using the arrow keys,
select date and press OK.
Note
Time format is 12 hour, and date format is day/month/year.
s
4-6
Type II Water System
Thermo Scientific
Section 4
Using Your TII
Timed Dispensing
Volumetric Dispensing
Reservoir
Replenishment
Setting the Displayed
Languages
Setting the Time and
Date