Your Thermo Scientific TII has been
shipped with one RO membrane for the
12 lph unit, or two RO membranes for
the 24 lph unit, which you will install.
The Startup Kit (D502138) includes a
prefilter, MPS cartridge, carbon filter, and
cartridge pack, and must be purchased
separately. To prepare your TII for
operation, disconnect the unit from the
power supply if connected.
Note
The cartridges must be installed in
the correct order to operate properly.
s
Particulates can damage your membranes,
resulting in premature membrane failure.
Therefore, your TII uses a prefilter to
remove particulates from your feed water.
Install the prefilter as follows:
1. Open the front left door.
2. Remove the prefilter (Part Number D502113) from its packaging and
wet the o-rings with water.
3. Insert the prefilter upright into the left position upper and lower quick
disconnects and push until firmly connected. For easier installation,
connect bottom first.
The MPS cartridge contains a slow dissolving anti-scalant material. The
anti-scalant material combines with water to prevent scale buildup on the
membrane surface. Install the MPS cartridge as follows:
1. Remove the MPS cartridge (Part Number D502114) from its
packaging and wet the o-rings with water.
2. Insert the MPS cartridge upright into the middle position upper and
lower quick disconnects and push until firmly connected. For easier
installation, connect bottom first.
Type II Water System
3-7
Thermo Scientific
Section 3
Installation
Cartridge / Membrane
Installation
Figure 3-8.
Cartridge
Installation:(left to right)
prefilter, MPS, carbon
MPS Cartridge
Installation