Specifications (cont.)
Table 2-4.
DI Product Water Quality (dispensed from unit)
Table 2-5.
Dimensions and Clearance Requirements
Table 2-6.
Environmental Conditions
Resistivity
ASTM Type II, >15 Megohm-cm
TOC
<15 PPB
Flow Rate
Up to 1 LPM with a new D3750 final filter
Bacteria
Less than 0.01 CFU/ml
Volumetric Dispense
Accuracy: ±5%, Repeatability: ±3%
Technical
Specifications
TII Dimensions
13.5" W X 18.5" D X 19.5" H (34.3 cm X 47.0 cm X 49.5 cm)
Clearances
Sides - 9" (22.9 cm) minimum to allow for servicing
Above -3" (7.6 cm) minimum for removal of top cover
Front - 4.75" (12.1 cm) minimum for opening front door
Electrical Requirements
Thermo Scientific TII is equipped with four power cords and cor-
responding fuses taped to each power cord to be plugged into a
grounded electrical outlet of the appropriate voltage.
All models: 90-240 VAC, +5% -10%, 47-63 Hz.
Operating
15°C -30°C; 20% -80% RH, non-condensing.
Installation Category II (over-voltage) in accordance with IEC 664.
Pollution Degree 2 in accordance with IEC 664.
Altitude limit
6562 ft (2,000 meters)
Storage
-25°C -65°C; 10% to 85% relative humidity