If an alternate reservoir is being used, one of three float/communication
configurations need to be purchased to connect the tank to the TII system.
Note
An alternative reservoir must have at least two ports to connect it to
the TII to allow for recirculation.
s
1. AY1403X2 – Float Assembly
Complete float assembly to be used with tank without a float. Contains
two floats and all connections required for communication with TII sys-
tem.
2. AY1403X4 – Float Communication Cable
Connects two user supplied floats to the TII system.
3. AY1403X5 – Jones Plug Communication Cable.
For use with tanks with Jones plug connections for level communication.
This includes Barnstead 60L reservoir, D2622 or 100L reservoir, D9021.
For detailed instructions on connecting the TII system to an alternate
tank, see the float or cable configuration instruction manual.
Note
TII system must be programmed to work with the alternate tank
configuration. Refer to “Reservoir Type” in the “System Options” section
of this manual to correctly configure the TII system.
Note
Test the float switch assembly in the reservoir to make sure the water
level registers on the display before you begin the sanitization mode. You
can do this by gently moving the float switches in the tank and looking to
the reservoir display to see if it registers. See Figure 3-20 to help locate the
reservoir display. This figure shows that the reservoir is 75% full.
Type II Water System
3-15
Thermo Scientific
Section 3
Installation
Reservoir Connection
(continued)
System Cleaning
Procedure
Reservoir Display
DI Display
RO Display
Figure 3-20.
Typical TII Display