Operation
Settings
Thermo Scientific
42iQD Instruction Manual
3-41
The Reaction Chamber screen provides status and alarms related to the
reaction chamber module. If an item being monitored goes outside the
lower or higher alarm limit, an alarm is activated.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Reaction
Chamber
The Reaction Chamber screen contains the following information:
●
Across:
●
Chamber:
This column lists items associated with the reaction
chamber.
●
Value:
Displays the current value for each item.
●
Low Alarm:
Displays low alarm status for each item.
●
High Alarm:
Displays high alarm status for each item.
●
Units:
Displays units for each item.
●
Down:
●
Instrument Temperature:
Displays the current instrument
temperature reading.
●
Flow:
Displays the current sample flow reading.
●
Chamber Pressure:
Displays the current chamber pressure reading.
●
Chamber Temperature:
Displays the current chamber temperature
reading. User can adjust low and high alarm limits.
●
Chamber Temp Sensor Short:
Displays OK/Fail for chamber
temperature sensor short.
Reaction Chamber