Operation
Data
Thermo Scientific
42iQD Instruction Manual
3-31
The Data Logging Setup screen allows the user to select data to be stored
and how it is stored.
Home Screen>Data>Advanced>Data Logging Setup
The Data Logging Setup screen contains the following information:
●
Select Data Logging Variables
: User selects instrument variables to log.
●
Period
: User selects how often data is collected by setting the duration
between logged data.
●
Data Treatment
: Toggles between Average, Current, Minimum and
Maximum. When set to average, the average value during the period
will be recorded. When set to current, the latest data will be recorded.
When set to minimum or maximum, the minimum or maximum value
during the period will be recorded.
●
Erase Data Log Records
: Allows the user to erase all values in the data log
and updates the data logging table.
Data Logging Setup