Troubleshooting
Troubleshooting Guide
Thermo Scientific
42iQD Instruction Manual
6-3
Problem
Possible Cause
Action
Defective cooler
Check temperature (less than -2
°
C at
T
amb
= 25
°
C).
Non-linear response
Incorrect calibration
source
Verify accuracy of multipoint
calibration source gas.
Leak in sample probe
line
Check for variable dilution.
Excessive response time
Partially blocked
sample capillary
Unplug power cord. Clean or replace
capillary.
Hang up/blockage in
sample filter
Change element.
Improper converter
operation
Questionable
calibration gas
Verify accuracy.
Converter
temperature too high
or too low
Temperature should be approximately
325
°
C.
Low line voltage
Check to see if line voltage is within
specifications.
Molybdenum
consumed
Replace Molybdenum converter
cartridge.
Alarm – Instrument
Temperature
Fan failure
Replace fan if not operating properly.
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting the
error and replace if needed.
Alarm – Bench
Temperature
Heaters failed
Replace heaters as needed.
Defective PCB
Replace PCB.
Alarm – Cooler Temp
Sensor
Check fan operation
Replace defective fan.
Check fan filter
Clean or replace filter.
Bad thermistor
Replace thermistor.
Alarm – NO
2
Converter
Temperature
Converter
temperature low
Molybdenum converter should be hot
to the touch. If not, the heater may
have failed. Check that converter
temperature set point is
approximately 325
°
C.
Alarm – Pressure
High pressure
indication
Check plumbing for leaks.
Check the pump for a tear in the
diaphragm.
Replace if needed.