Maintenance
Capillaries Inspection and Replacement
Thermo Scientific
42iQD Instruction Manual
5-3
The capillaries normally only require inspection when instrument
performance indicates that there may be a flow problem.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
▲
Use the following procedure to inspect and replace the capillaries. This
procedure can be used to check any or all of the capillaries.
1.
Turn instrument OFF, unplug power cord and remove the cover
).
Figure 5–1.
Inspecting and Replacing the Capillaries
Capillaries
Inspection and
Replacement