Maintenance
Thermoelectric Cooler Fins Inspection and Cleaning
5-2
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to inspect and clean the thermoelectric cooler
fins.
1.
Turn the instrument OFF and unplug the power cord.
2.
Remove the PMT cooler from the instrument. Refer to “
” on page
3.
Blow off the cooler fins using clean pressurized air. It may be more
convenient to vacuum the cooler fins. In either case, make sure that
particulate accumulation between the fins has been removed.
4.
If necessary, use a small brush to remove residual particulate
accumulation.
Thermoelectric Cooler
Fins Inspection and
Cleaning