Operation procedure
8. Press the green-blinking V key.
The light then goes out.
Input the external analog signal into the connector
on the analog IF board.
The voltage indicator displays the set value that
corresponds to the external analog signal.
The display range of the set voltage is from -10%
to +110% of the rated voltage.
The analog IF board connector varies according
to the type of external analog signal.
◆
For details,
“9-1-2. Wiring an analog signal to
and
The voltage that can be set by performing this operation may exceed the maximum rated voltage of the unit, depending
on how it is used.
When the voltage is set lower than 0V or higher than 102.5% of the rated voltage, all the digits on the voltage indicator
will blink, and the output will be turned from on to off.