Status
Condition
Action
Activate
0x0073 to
MACSubcmd()
Outputs 24 bytes of IT data values on
MACData()
in the following
format: aaAAbbBBccCCddDDeeEEffFFggGGhhHHiiIIjjJJkkKKllLL
where:
AAaa: True Rem Q. True remaining capacity in mAh from IT
simulation before any filtering or smoothing function. This value can
be negative or higher than FCC.
BBbb: True Rem E. True remaining energy in cWh from IT simulation
before any filtering or smoothing function. This value can be
negative or higher than FCC.
CCcc: Initial Q. Initial capacity calculated from IT simulation
DDdd: Initial E. Initial energy calculated from IT simulation
EEee: TrueFullChgQ. True full charge capacity
FFff: TrueFullChgE. True full charge energy
GGgg: T_sim. Temperature during the last simulation run (in 0.1K)
HHhh: T_ambient. Current estimated ambient temperature used by
the IT algorithm for thermal modeling
IIii: RaScale 0. Ra table scaling factor of Cell1
JJjj: RaScale 1. Ra table scaling factor of Cell2
KKkk: CompRes1. Last computed resistance for Cell1
LLll: CompRes2. Last computed resistance for Cell2
12.2.40 MACSubcmd() 0x0074 ITStatus2
This command instructs the device to return Impedance Track related gauging information on
MACData()
.
Status
Condition
Action
Activate
0x0074 to
MACSubcmd()
Outputs 24 bytes of IT data values
on
MACData()
in the following format:
AABBCCDDEEFFggGGhhHHiiIIjjJJkkKKllLLmmMMnnNNooOO
where:
AA: Pack Grid. Active pack grid point (only valid in discharge)
BB: LStatus—Learned status of resistance table
Bit 3 | Bit 2 | Bit 1| Bit 0
QMax | ITEN | CF1 | CF0
CF1, CF0: QMax Status
0,0 = Battery OK
0,1 = QMax is first updated in learning cycle.
1,0 = QMax and resistance table updated in learning cycle
ITEN: IT enable
0 = IT disabled
1 = IT enabled
QMax: QMax field updates
0 = QMax is not updated in the field.
1= QMax is updated in the field.
CC: Cell Grid 1. Active grid point of Cell1
DD: Cell Grid 2. Active grid point of Cell2
EE: N/A. RSVD
FF: N/A. RSVD
HHhhGGgg: State Time. Time passed since last state change
(Discharge, Charge, Rest)
IIii: DOD0_1. Depth of discharge for Cell1
JJjj: DOD0_2. Depth of discharge for Cell2
KKkk: DOD0 Passed Q. Passed charge since DOD0
LLll: DOD0 Passed Energy. Passed energy since the last DOD0
update
MMmm: DOD0 Time. Time passed since the last DOD0 update
NNnn: DODEOC_1. Cell 1 DOD@EOC
OOoo: DODEOC_2. Cell 2 DOD@EOC
12.2.41 MACSubcmd() 0x0075 ITStatus3
This command instructs the device to return Impedance Track related gauging information on
MACData()
.
I2C Commands
90
BQ28Z610-R2
Copyright © 2022 Texas Instruments Incorporated