4 Transmitter
48
4.5.2.
Warning messages
Warning messages show an early warning or a current malfunction
which may negatively impact measuring.
Message Display
Cause
Remedying of fault
00E00
T ambient high The ambient
temperature exceeds
the permissible
temperature for the
transmitter
Take necessary
measures to lower
ambient temperature,
e.g. through venting or
cooling
00E01
T ambient low
The ambient
temperature is below
the permissible
temperature for the
transmitter
Take necessary
measures to raise
ambient temperature,
e.g. through heating
00809
Pressure too
high
The process pressure
exceeds the pressure
intended for the
transmitter
Remove the transmitter
from the process and
take any necessary
measures to lower the
pressure
00E02/00E
05
Supply voltage
low
The supply voltage is
below the minimum
voltage required for the
transmitter
Take measures to
ensure sufficient
voltage supply
0081C
Alarm 1
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081D
Alarm 2
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081E
Alarm 3
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081F
Alarm 4
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
Summary of Contents for 6351
Page 2: ...2 ...
Page 33: ...4 Transmitter 33 4 4 5 Overview of the testo 6351 user menu ...
Page 34: ...4 Transmitter 34 ...
Page 35: ...4 Transmitter 35 ...
Page 84: ...0970 6353 en 03 ...