4 Transmitter
42
1
Channel 1
test contacts
2
Service
interface
3
Multimeter
4.
Return to
Test Analog Output
with
ESC
and use
or
to
continue to
Test Relay Output
.
Testing functionality of the pressure sensor (
Test pressure
sensor
)
This function is only required to calibrate the pressure
sensor.
Testing functionality of relay outputs
1.
Press
SET
and choose between
Alarm 1, 2, 3, 4
with
or
.
2.
Press
SET
.
The relay can now be tested. You can choose between
OFF
and
ON
using
or
. If
ON
is chosen, the NO contact is
closed, the NC contact opened. If
OFF
is chosen, the NC
contact is closed, the NO contact opened.
3.
To test, route a measuring cable from the relay terminals () out
of the transmitter to a multimeter (resistance measurement) or
continuity tester.
4.
Return to
Test Relay Output
with
SET
(starts relay test) or
ESC
(exits the menu without relay test).
Reading off min./max. values of channels
To reset the max./min. values, see 4.4.6.8. Editing Reset main
menu.
1.
Read off the min./max. values of the channel in succession with
or
and return to the
Main Menu Analysis
with
ESC
.
2.
Continue to
Main Menu Message
with
or
or return to
Measuring Mode with
ESC
.
Summary of Contents for 6351