5 Parameterizing, adjusting and analyzing software (P2A software)
69
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark transmitter test.
- Current operating hours are shown.
3.
Confirm control query to perform the reset.
- The values are reset to the customer-specific factory settings.
4.
Click on
[OK]
or
[Cancel]
to close the dialogue.
5.3.3.3.
Testing analog output
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark channel and test values.
Field/button
Explanation
Transmitter test
Check the analog outputs (see
4.4.6.4. Testing functionality of analog
outputs).
Current reading
Readings are updated every second.
Unit
Unit according to the type of analog
output.