4 Transmitter
41
4.
Return to
Pressure process data
with
ESC
.
5.
Return to
Change parameters
with
ESC
and use
or
to
continue to
Code
.
Editing code settings
You can set the access code (password).
If a code other than "0000" (factory setting) is set, the
transmitter can only be operated once this code has been
entered via the menu.
1.
Select
Code
with
or
and confirm selection with
SET
.
2.
Scroll one digit to the right using
and increase value of digit
by 1 using
. Confirm with
SET
or abort entry with
ESC
.
3.
Return to
Code
with
ESC
.
4.
Return to
Main Menu Settings
with
ESC
.
4.4.6.4.
Editing Main Menu Analysis
You can test the functionality of analog and relay outputs. In
addition, you can read off the minimum and maximum values (since
the last voltage supply or reset of the min./max. values).
Testing functionality of analog outputs
This function affects the analog outputs directly, not only
the test contacts.
1.
In the Measuring Mode press
SET
, select
Main Menu Analysis
with
or
and confirm selection with
SET
.
-
Test Analog Output
is shown.
2.
Press
SET
, scroll one digit to the right using
and increase
value of digit by 1 using
. Any analog output value can be
predefined, e.g. for an analog output of 4 to 20 mA, the value
"6.0 mA". Confirm with
SET
or abort entry with
ESC
.
3.
Accept setting by pressing
SET
and test with multimeter
(minimum requirement: resolution 6.5 digits, at least 2-times
more accurate than the 6351) as follows:
Analog output 1: Via test contacts under service flap, see
diagram.