4 Transmitter
49
4.5.3.
Transmitter error messages
Error messages show a current malfunction.
Message Display
Cause
Remedying of fault
01505
Watchdog error Due to a processor
error, the transmitter
performs an automatic
restart.
If the problem occurs
frequently, contact
Testo Service.
01115
Low adjustment
temperature
The ambient
temperature is too low
during the pressure
adjustment.
Take necessary
measures to raise
ambient temperature,
e.g. through heating.
01116
High adjustment
temperature
The ambient
temperature is too high
during the pressure
adjustment.
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
4.5.4.
Handling alarm messages
Shown on the
display
5
Can be used for
collective alarm
6
Additional
message end
7
New limit value
x
Scaling changed
x
Adjustment DeltaP
x
Pressure too high
x
x
Alarm 1
x
Alarm 2
x
Alarm 3
x
Alarm 4
x
Transmitter reset
x
5
If multiple messages/alarms are activated at the same time, only the last message/alarm is shown. If this is
cancelled, the other messages that are still active are no longer displayed.
6
The message can be assigned the collective alarm function, which means that the collective alarm is activated
as soon as at least one of the messages assigned to it is activated. The collective alarm can be assigned to each
of the 4 optional relays. The collective alarm is then always the same, as it can only be defined once.
7
The message is shown upon the occurrence of the event causing the message as well as when closing. Two
entries appear in the history in the P2A software: Message text_start and Message text_end.