5 Parameterizing, adjusting and analyzing software (P2A software)
71
5.3.3.4.
Testing switch output relays 1 to 4
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark
Relay tests
and test the values.
Field/button
Explanation
Test the relay function (see 4.4.6.4. Testing
functionality of relay outputs).
[Activate relay n]
Close contact.
A warning informs that the value is being
transmitted to a connected PLC, external
display, etc. in the event of existing cabling.
[Deactivate relay
n]
Open contact.
A warning informs that the value is being
transmitted to a connected PLC, external
display, etc. in the event of existing cabling.
3.
Click on
[OK]
or
[Cancel]
to close the dialogue.
- The analog output returns to Measuring Mode again.