4 Transmitter
45
3.
Use
or
to select
Adj. Point 1
.
4.
Press
SET
. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using
and increasing the value of digit by 1 using
. Confirm with
SET
or abort entry with
ESC
.
5.
Use
or
to select
Adj. Point 2
.
6.
Press
SET
. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using
and increasing the value of digit by 1 using
. Confirm with
SET
or abort entry with
ESC
.
7.
Use
or
to select
Adj. Point 3
.
8.
Press
SET
. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using
and increasing the value of digit by 1 using
. Confirm with
SET
or abort entry with
ESC
.
9.
Return to main menu
Adjust
with
ESC
.
10.
Continue to main menu
Reset
with
or
or return to
Measuring Mode with
ESC
.
Performing the pressure adjustment
Also see 4.3.3.2. n-point adjustment, for instructions on how to
perform the pressure adjustment.
1.
In the Measuring Mode press
SET
, select main menu
Adjust
with
or
and confirm selection with
SET
.
The pressure adjustment can be performed at 3 to 6
adjustment points.
2.
Select
Adjust Pressure
with
or
and confirm with
SET
.
3.
Use
or
to select
Adj. Point 1
.
4.
Press
SET
. Read off pressure sensor display and enter this
value in the user menu. Do this by scrolling one digit to the right
using
and increasing the value of digit by 1 using
. Confirm
with
SET
or abort entry with
ESC
.
5.
Repeat steps 1 to 4 for adjustment points 2 to 6.
6.
Return to main menu
Adjust
with
ESC
.
7.
Continue to main menu
Reset
with
or
or return to
Measuring Mode with
ESC
.