Signal Analyzer
117
4.
Record the measurement of the Frequency Counter and compare with the specification.
Item
Measurement Result [MHz]
Frequency Counter Measurement Value
19.2.2 Input Power Level Accuracy
19.2.2.1 CW Mode Power Level Accuracy
Specification
: ±1 dB
Objective
: To measure the level accuracy of the input CW signal.
1.
Test Setup: Connect the device as shown below.
2.
Configure MTP200B as shown below.
A. FCN -> Recall -> RESET <Resets the device>
B. MODE: ANL -> MODE -> CW <Set the measurement mode CW>
C. FREQUENCY: Set the measurement frequency to 2412.000 MHz.
3.
VSG(Vector Signal Generator) Setting:
A. RESET <Resets the device>
B. Amplitude: +10 dBm <Sets the output power>
C. RF: ON <Turns on the output power>
4.
Connect the RF output of VSG to MPT200A RF input.
5.
Synchronize the time base of the two devices by connecting 10 MHz REF Out of VSG to MTP20
0B 10 MHz REF In.
6.
Set the frequency/channel of VSG and MTP200B as shown below and record the “INPUT_POWE
R” displayed on MTP200B.
Summary of Contents for MTP200B
Page 1: ...MTP200B WLAN BT LE Tester Operating Manual R20160511 ...
Page 10: ...x ...
Page 11: ...xi ...
Page 12: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 30: ...14 GEN Generator Menu Tree 4 2 GEN Generator Menu Tree Figure 4 2 GEN Generator Menu Tree ...
Page 31: ...ANL Analyzer Menu Tree 15 4 3 ANL Analyzer Menu Tree Figure 4 3 ANL Analyzer Menu Tree ...
Page 34: ......
Page 41: ...PART 2 WLAN Test 25 ...
Page 42: ......
Page 62: ......
Page 63: ...PART 3 BT LE Test 47 ...
Page 64: ......
Page 68: ......
Page 76: ......
Page 82: ......
Page 83: ...PART 4 Programming Guide 67 ...
Page 84: ......
Page 102: ......
Page 103: ...PART 5 Command List 87 ...
Page 104: ......
Page 129: ...PART 6 User s Device Diagnosis 113 ...
Page 130: ......
Page 144: ......
Page 145: ...PART 7 Appendix 129 ...
Page 146: ......
Page 156: ......