116
Signal Analyzer
19.1.2 Default Setting of MTP200B
Before starting measurement, configure MTP200B as described below. In GEN mode, a damag
ed cable will be considered missing cable (Path Loss = 0).
1.
Turn on the power.
2.
Fully pre-heat the device.
3.
Press the
(
FCN)
->
(SAVE)
keys and select “RESET” to result the device.
4.
Press
(
M1)
to shift to GEN mode.
19.2 Signal Analyzer
19.2.1 Frequency Accuracy
Specification
: 2400.00 MHz ± 1 ppm
Objective
: To measure the frequency deviation of the reference signal
1.
Test Setup: Connect the device as shown below.
2.
MTP200B Setting
A. FCN -> Recall -> RESET <Resets the device>
B. MODE: GEN -> CW <Set the measurement mode CW>
C. FREQUENCY: Set the measurement frequency to 2400.00 MHz.
D. POWER: -5 dBm
E. RUN: ON
3.
Configure the frequency Counter as follows:
A. Press
Freq Ch 2
key
Summary of Contents for MTP200B
Page 1: ...MTP200B WLAN BT LE Tester Operating Manual R20160511 ...
Page 10: ...x ...
Page 11: ...xi ...
Page 12: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 30: ...14 GEN Generator Menu Tree 4 2 GEN Generator Menu Tree Figure 4 2 GEN Generator Menu Tree ...
Page 31: ...ANL Analyzer Menu Tree 15 4 3 ANL Analyzer Menu Tree Figure 4 3 ANL Analyzer Menu Tree ...
Page 34: ......
Page 41: ...PART 2 WLAN Test 25 ...
Page 42: ......
Page 62: ......
Page 63: ...PART 3 BT LE Test 47 ...
Page 64: ......
Page 68: ......
Page 76: ......
Page 82: ......
Page 83: ...PART 4 Programming Guide 67 ...
Page 84: ......
Page 102: ......
Page 103: ...PART 5 Command List 87 ...
Page 104: ......
Page 129: ...PART 6 User s Device Diagnosis 113 ...
Page 130: ......
Page 144: ......
Page 145: ...PART 7 Appendix 129 ...
Page 146: ......
Page 156: ......