10. BT LE Signal Analyzer
General Configuration for BLE TX Measurement
53
MTP200B sends user-defined BT LE packets to DUT while DUT checks the number of received packets to measure the
sensitivity and the Packet Error Rate (PER).
10.1 General Configuration for BLE TX Measurement
Figure 10-1
Configuration for BLE TX Measurement Using a Shield Box
1. DUT Preparation:
Most BLE terminals have a test mode for performance checking. To set the test
mode, users use the PC control program or perform the required terminal operations depending o
n DUT. In test mode, set the BLE channel, packet length, and data type to be measured and send T
X signals.
2. Shield Box:
For the reliable measurement and prevention of electromagnetic interferences, use a s
hield box. Connect the RF cable to the shield box or create a radiation environment using the ante
nna coupler depending on the DUT type.
3. MTP200B:
MTP200B checks the measurement result on the front LCD or carries out measurement
using a separate user remote program. Production lines that require fast measurement tend to use
the remote program to control DUT and MTP200B.
Summary of Contents for MTP200B
Page 1: ...MTP200B WLAN BT LE Tester Operating Manual R20160511 ...
Page 10: ...x ...
Page 11: ...xi ...
Page 12: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 30: ...14 GEN Generator Menu Tree 4 2 GEN Generator Menu Tree Figure 4 2 GEN Generator Menu Tree ...
Page 31: ...ANL Analyzer Menu Tree 15 4 3 ANL Analyzer Menu Tree Figure 4 3 ANL Analyzer Menu Tree ...
Page 34: ......
Page 41: ...PART 2 WLAN Test 25 ...
Page 42: ......
Page 62: ......
Page 63: ...PART 3 BT LE Test 47 ...
Page 64: ......
Page 68: ......
Page 76: ......
Page 82: ......
Page 83: ...PART 4 Programming Guide 67 ...
Page 84: ......
Page 102: ......
Page 103: ...PART 5 Command List 87 ...
Page 104: ......
Page 129: ...PART 6 User s Device Diagnosis 113 ...
Page 130: ......
Page 144: ......
Page 145: ...PART 7 Appendix 129 ...
Page 146: ......
Page 156: ......