Signal Generator
125
4.
Set the frequency and power of MTP200B as shown below and record the measurement by the
spectrum analyzer.
5.
Measurement results
Table 19-9
BLE Power Generator Level Accuracy Test (Spectrum Analyzer)
Level
Frequency (CH)
2402 MHz (0 CH)
2440 MHz (19 CH)
2480 MHz (39 CH)
-5 dBm
-20 dBm
-40 dBm
-60 dBm
-70 dBm
19.3.3.2 Measurement Procedure using Power Meter
1.
Test Setup: Connect the device as shown below.
2.
MTP200B Setting:
A. FCN -> Recall -> RESET <Resets the device>
B. MODE: GEN -> MODE -> BT_LE <Select the BT LE measurement mode>
C. CH_NUM: 0 <Select channel number 0>
D. BIT_PATTERN : PRBS9
E. LENGTH: 37 B
F. POWER : -5dBm
G. RUN: ON
3.
POWER METER Setting:
A. RESET <Resets the device>
4.
Set the frequency and power of MTP200B as shown below and record the measurement by the
Power Meter.
Summary of Contents for MTP200B
Page 1: ...MTP200B WLAN BT LE Tester Operating Manual R20160511 ...
Page 10: ...x ...
Page 11: ...xi ...
Page 12: ......
Page 14: ......
Page 16: ......
Page 18: ......
Page 30: ...14 GEN Generator Menu Tree 4 2 GEN Generator Menu Tree Figure 4 2 GEN Generator Menu Tree ...
Page 31: ...ANL Analyzer Menu Tree 15 4 3 ANL Analyzer Menu Tree Figure 4 3 ANL Analyzer Menu Tree ...
Page 34: ......
Page 41: ...PART 2 WLAN Test 25 ...
Page 42: ......
Page 62: ......
Page 63: ...PART 3 BT LE Test 47 ...
Page 64: ......
Page 68: ......
Page 76: ......
Page 82: ......
Page 83: ...PART 4 Programming Guide 67 ...
Page 84: ......
Page 102: ......
Page 103: ...PART 5 Command List 87 ...
Page 104: ......
Page 129: ...PART 6 User s Device Diagnosis 113 ...
Page 130: ......
Page 144: ......
Page 145: ...PART 7 Appendix 129 ...
Page 146: ......
Page 156: ......