Page
51
of
76
Low flow time
0
–
60s
Default: 15s
Time in seconds the low flow rate pre-alarm is activated. When the flow rate is
still too low after this time, the low flow rate alarm will be activated. When set to
0, the pre-alarm is disabled and the low flow rate alarm will be activated
immediately in case of a low flow rate.
High flow limit
10
–
300ml/min
Default: 200ml/min
When the flow rate passes above this limit during donation, the clamp will close
and the high flow rate alarm will be triggered.
Cont high flow
Yes / No
Default: No
Yes:
After a high flow alarm, the user will be able to continue the ongoing
donation.
No:
After a high flow alarm, the ongoing donation will be discontinued.
Max. don. time
00:01
–
30:00 (mm:ss)
Default: 15:00
Sets the maximum donation time. When the elapsed time passes this limit, the
donation is stopped automatically, the clamp is closed and an alarm is triggered.
Warning timer
00:00
–
30:00 (mm:ss)
Default: 12:00
A special time warning can be given on the touch screen using this setting. When
the elapsed time passes the limit, the
’time elapsed’
as shown on the touch screen
turns red.
When the estimated time to finish the donation passes the limit, the
‘
estimated
time
left’
turns red.
Using this setting, it can be visualized during donation whether or not the blood
will be usable for making platelet concentrate. When setting the timer to 00:00,
the warning timer is disabled.
Time estim.
Left / Tot
Default: Left
Left:
During donation, the estimated time before finishing the donation is
displayed. The value is calculated based on the flow rate, the total
donation volume and the time elapsed.
Tot:
The estimated total donation time is shown.
Enable pause
Yes / No
Default: No
Yes:
Special option for labeling of the bag during donation. Between 1%
and 80% of the donation (based on donated volume), a pause button
is displayed. When pressed, the clamp will close and the user has 60
seconds to put labels on the blood bag. After that time, or when
pressing ‘resume’, the donation continues automatically. The pause
function is only available once during a donation.
No:
The pause option is not available. A donation can still be paused by
pressing ‘S
top
’ and ‘Continue’.
Summary of Contents for T-RAC II