UM-BIOIIADVANCE-EN-1302
31
10.
TROUBLESHOOTING
No.
Type of
error
Causes of error
Correction of error
1
Insufficient
exhaust
flow
Alarm “
LOW
EXH. FLOW”
(see Chapter
4.5 point D)
Exhaust partially covered.
Visually check the exhaust to remove any
type of obstruction.
Front perforation in the
working area
totally/partially covered.
Visually check the perforation and remove
any type of obstruction.
Distance between cabinet
exhaust and lab ceiling too
small.
Increase space between cabinet exhaust
and the ceiling. In some cases it is
possible to modify or eliminate the false
ceiling.
Install a lower support table. This means
the working area will be also lower.
Duct the cabinet (
Chapter 3.5
).
Clogged exhaust filter.
Check filter clogging with the cabinet in
normal operation (
Chapter 4.5 point 2
).
Check the date of the last filter change
and the hours it has been under operation.
(
Chapter 4.5 point 3.3.1
).
Notify the technical service to change the
filter.
Probe not working properly. Check the probe position is correct.
Ducted cabinet and ducting
too long.
Fit an additional fan at the end of the duct
(
Chapter 3.5
).
2
Insufficient
laminar
down flow
Alarm “
LOW
LAM. FLOW
”
(see Chapter
4.5 point D)
Front and rear perforation
in the work zone
totally/partially covered.
Visually check the perforation and remove
any type of obstruction.
Clogged laminar down flow
filter.
Check filter clogging with the cabinet in
normal operation (
Chapter 4.5 point 2
).
Check the date of the last filter change
and the hours it has been under operation.
(
Chapter 4.5 point 3.3.1
).
Notify the technical service to change the
filter.
Probe not working properly.
Check the probe position is correct (sensor
parallel to the flow).
Check and remove any obstruction to the
probe.
Summary of Contents for Bio II Advance
Page 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Page 3: ...UM BIOIIADVANCE SP 1302 2...
Page 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Page 37: ...UM BIOIIADVANCE EN 1302 2...
Page 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Page 71: ...UM BIOIIADVANCE FR 1302 2...
Page 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Page 105: ...UM BIOIIADVANCE DE 1302 2...
Page 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Page 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Page 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Page 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Page 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Page 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Page 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Page 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Page 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Page 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Page 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Page 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Page 226: ......
Page 227: ......