T3PSX3200P Series User Manual
194
Parallel
regulation
Line:
≤
0.01% + 3mV
Load:
≤
0.01% +3mV(rating current
≤
3A)
≤
0.02% +5mV(rating current > 3A)
Series
regulation
Line:
≤
0.01% + 5mV
Load:
≤
100mV
Ripple & noise
≤
1mVrms(5Hz-1MHz)
Resolution
Voltage
Current
Voltage: programming 1mV ,readback
0.1mV
Current: 0.1mA (T3PS13206: 0.2mA)
Accuracy
Ammeter
3.2A full scale (T3PS13206: 6.2A),
programming 5 digits, readback 5 digits
Voltmeter
33V full scale, programming 5 digits,
readback 6 digits
Setting
accuracy
Voltage: ± (0.03% of r 10mV)
Current: ± (0.3% of r 10mA)
Readback
accuracy
Voltage: ± (0.03% of r 10mV)
Current: ± (0.3% of r 10mA)
Bindpost
Output voltage
Output current
1.8V/2.5V/3.3V/5.0V, ±5%
5A
CH3
(T3PS33203P)
Line regulation
≤
3mV
Load regulation
≤
5mV
Ripple & noise
≤
2mVrms (5Hz - 1MHz)
Transient
recovery time
≤
100µs
(50% load change, minimum load 0.5A)
USB Port
Output
1.8V/2.5V/3.3V/5.0V, ±0.35V, 3A
Warning
The output current from the 2 terminals should Not
exceed 5A.
Load Mode
Display
Voltage
1-33.00V
Current
0-3.200A(T3PS13206:0-6.200A)
Power
0-50.00(T3PS13206:0-100.00W)
Summary of Contents for T3PS13206P
Page 1: ... ...
Page 106: ...T3PSX3200P Series User Manual 106 Register Commands OPC 187 OPC 187 ...
Page 177: ...REMOTE CONTROL 177 Example SAV 1 Recalls the setting stored in memory 2 STATE02 ...
Page 185: ...REMOTE CONTROL 185 Example STB Returns 81 if the status byte register is set to 0101 0001 ...
Page 196: ...T3PSX3200P Series User Manual 196 Between chassis and DC power cord 30MΩ or above DC 500V ...
Page 197: ... 0 0 0 1 2 3 4 5 6 7 8 8 9 3 3 3 45 6 990 0 0 9 0 0 0 9 0 0 0 0 9 7 7 88 8 1 931711 RevB ...