In general it is best to use the Operation Complete Bit (bit 0) in the
standard event register to check to see if an operation/command
has completed. After executing the *OPC command, the OPC bit
will be set to 1. If a command or query is placed in the output
buffer immediately before the *OPC command is sent, the
Operation Complete Bit can be used to determine when the
information can be used. However if too many commands/queries
are executed prior to the execution of the *OPC command, the
output buffer could become saturated and the unit will stop taking
readings.
Standard Event Register
The Standard Event Register reports the following types or events:
Power on has been detected, command syntax errors, command
execution errors, self test and execution errors, query errors or if the
*OPC command is executed. Any one or more of these events will
set the standard event summary bit in the status byte register. To
set a mask for the enable register, a binary-weighted decimal
number must be written using the *ESE command.