Errors
Error Message
Errors are stored in a first in-first out (FIFO)
order. The first error message that is returned is
the first error message that was stored. When
an error is read it is also cleared from the
queue.
If there are more than 10 errors produced the
last error in the queue is replaced with “Que
overflow”. Unless the error queue is cleared,
no more errors can be written to the error
queue. If there are no errors in the error queue,
the instrument will return “No error”.
To clear the error queue, you can use
the
:SYSTem:CLEar
command or cycle the
power. When you read a message from the
error queue that message will be cleared from
the error queue. Using the *RST command to
reset the instrument does not clear the error
queue.
Remote control instructions can be used to
clear the error queue. See the instructions listed
in the previous chapter for details.