REMOTE CONTROL
139
Example
:SEQUence2:CYCLEs N,20
Sets the cycle number of sequence as 20
.
:SEQUence[1|2]:CYCLEs?
Description
Queries the cycle number of the sequence.
Example
:SEQUence2:CYCLEs?
Queries the cycle number of the sequence.
Returns to I or N, <value>.
:SEQUence[1|2]:ENDState {OFF|LAST}
Description
Sets terminate state of sequence.
Parameter
OFF
Output disabled (OFF): instrument
automatically disable output when
output is complete.
LAST
The last group (LAST): instrument stays
at the output state of the last group
when output is complete.
Example
:SEQUence2:ENDState LAST
Sets the terminate state of sequence as “the last
group”.
:SEQUence[1|2]:ENDState?
Description
Queries the terminate state of the sequence.
Example
:SEQUence2:ENDState?
Queries the CH2terminate state of the sequence.
Returns to OFF or LAST.