![Teledyne Princeton Instruments PI-MAX 3 System Manual Download Page 89](http://html.mh-extra.com/html/teledyne/princeton-instruments-pi-max-3/princeton-instruments-pi-max-3_system-manual_1080556089.webp)
Chapter 7
Operation
89
7.13 WinX/32 Experiment Setup
This section provides information about configuring each system parameter within
WinX/32.
7.13.1
Experiment Setup
►
Main Tab
In WinX, the first experiment setup parameters are typically configured on the
Experiment Setup
►
Main tab. See
Figure 7-17: Typical WinX/32 Experiment Setup
►
Main Tab Dialog
These parameters determine such things as the number of separate frames (denoted as
images or spectra) that will be acquired, how much of the CCD will be used to acquired
data, the intensifier gain, and the intensifier mode. The Main parameters are described
briefly below.
•
Number of Images/Spectra and Accumulations
This is the number of sequential frames the software will collect and store in a
single file. Each frame will be digitized and stored, but all of the frames will be in
a single file. A frame may contain data from more than one region of interest as
defined via the ROI Setup tab.
If the Accumulations > 1, each stored frame will contain summed data sets from
two or more exposures. The number of separate frames stored will the same as
it would be for Accumulations =1, but the total number of exposures required
would be the product of the Number of Images/Spectra parameter times the
Accumulations value.
•
CCD Readout
This parameter selection determines whether the entire CCD array will be used
to acquire a frame or if only a rectangular portion (or portions) of the array will
be used. When Use Region of Interest is selected, data will be collected from the
region(s) defined on ROI Setup tab. The full chip size or the sum of the active
region(s) as defined on the ROI tab is displayed.
4
4
11-
01
29
_
0018
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...