![Teledyne Princeton Instruments PI-MAX 3 System Manual Download Page 88](http://html.mh-extra.com/html/teledyne/princeton-instruments-pi-max-3/princeton-instruments-pi-max-3_system-manual_1080556088.webp)
88
PI-MAX
®
3 System Manual
Issue 2
•
Image Shift {Shifting Under Mask}
This level is at a logic high while the image is being shifted under the mask.
•
Logic 1 {Always High}
The level at the connector is high.
•
Read Out {Reading Out}
It is at a logic high when CCD is being read; otherwise low.
•
Shutter {Shutter Open}
This level is at a logic high when the photocathode is gated on (signal is
accumulating on the CCD) and can be used to drive an external
electro-mechanical shutter. Contact the factory for more information.
•
Wait for Trigger {Waiting For Trigger}
This level is at a logic high when the camera is ready to acquire and is waiting
for an external trigger (through the EXT SYNC connector) before exposing the
CCD. The level goes low when a trigger is detected: exposure begins. The Wait
for Trigger (WFT) signal goes high immediately after readout or after preopen (if
it is active). If continuous cleans is enabled, the camera will check for a trigger at
the EXT SYNC connector before entering a continuous clean cycle. If none has
occurred, a cleaning cycle is initiated and completed. Before the next cycle
begins, the EXT SYNC connector is checked again, and exposure will start if a
trigger has occurred.
When the Invert LOGIC {Invert Output Signal} check box is checked, the output is at a
logic low when the action is occurring.
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...