![Teledyne Princeton Instruments PI-MAX 3 System Manual Download Page 68](http://html.mh-extra.com/html/teledyne/princeton-instruments-pi-max-3/princeton-instruments-pi-max-3_system-manual_1080556068.webp)
68
PI-MAX
®
3 System Manual
Issue 2
Figure 7-3: Typical WinX/32 Hardware Setup Dialog: Phosphor Decay Delay
Figure 7-4: Typical LightField Common Acquisition Settings
►
Advanced
►
Phosphor
Decay Delay Dialog
7.5
Temperature Control
CAUTION!
!
Under normal conditions, the front end of the camera is
sealed and backfilled so there is no danger of damage due
to condensation.
7.5.1
Cooling Method
Cooling the CCD array reduces the amount of dark charge thereby acting to improve
the signal-to-noise ratio. An internal Peltier device directly cools the cold finger on
which the CCD is mounted. The heat produced by the Peltier device is then removed by
the air drawn into the camera by the internal fans and exhausted through the side-panel
grill. The fans are always running and cooling the CCD and the internal electronics.
Additional heat removal can be performed by circulating coolant through an internal
heat block.
4
4
11
-0
129
_
003
0
E
NTER
D
ESIRED
P
HOSPHOR
D
ECAY
D
ELAY
T
IME
44
11
-0
12
9_
0
0
31
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...